Measurements have been made on the computerized transverse axial scanning equipment to estimate the output of the X-ray generator, to investigate the radiation field in the X-ray room and to estimate dose to the patient. Isodose curves are shown for the total radiation field from the machine, for that fraction of the field which is due to tube housing leakage and for the fraction due to scatter from the main beam. Measurements in a simple head phantom, and subsequent calculations, showed that the exposure from a series of complete transverse axial scans varies between 1·0 and 2·5 R over the cranium. Estimates indicate that the integral dose from a single complete scan is likely to be less than that from a single skull X ray. Gonad dose is considerably less than 0·1 mrad for each complete scan.