Answers are grounded in this paper and cite their sources.
BACKGROUND: Mechanical closure of percutaneous femoral arteriotomies following catheter based procedures remains problematic. METHODS: The EVS closure device is the first to utilize a staple to effect arteriotomy closure and was compared to manual compression following sheath removal in a 362 patient randomized (2:1 to device) multicenter trial. As pre-specified, one half of the patients underwent coronary intervention. RESULTS: Time to hemostasis was significantly reduced in the EVS group for both diagnostic (3.3 +/- 2.6 vs. 19.3 +/- 5.7 minutes; p < 0.001) and interventional procedures (5.5 +/- 5.1 vs. 22.3 +/- 9.9 minutes; p < 0.0001). Time to ambulation was similarly reduced in the EVS group following diagnostic (2.4 +/- 3.3 vs. 6.0 +/- 5.2 hours; p < 0.001) and interventional procedures (3.4 +/- 4.5 vs. 7.6 +/- 7.0 hours; p < 0.001). The incidence of major complications was similar between the EVS and manual compression groups at discharge (0.4% vs. 1.7%; p = NS) and at 30 day follow-up (0.4% vs. 2.5%; p = NS). CONCLUSION: Compared to manual compression, the EVS device provides a safe and effective method of femoral artery closure.
No takes yet. Share an insight, caveat, or question.
Ansel et al. (2006) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: