Key result
Fabry-Perot leaky-wave antenna successfully detects BR and HR at up to 3 meters.
Why the study?
Antenna requirements needed to be addressed for continuous remote vital sign monitoring of a patient during sleep in a dynamic environment using Doppler radar.
A novel 60 GHz Fabry-Perot beam scanning antenna enables accurate remote monitoring of breathing and heart rates up to 3 meters away, allowing for patient movement.
May support contactless vital-sign monitoring in research settings; leaves open prospective clinical validation before practice adoption.
A beam scanning Fabry-Perot leaky-wave antenna (LWA) is designed at 60 GHz-band for single tone continuous-wave Doppler radar (DR)-based remote vital sign monitoring (RVSM) in a dynamic environment. A brief analysis is presented on antenna requirements for the DR in a typical dynamic scenario where a continuous RVSM is desired of a patient during sleep. The antenna is designed based on the drawn requirements with sufficiently high gain (>18 dBi) and wide beam scanning (12°) over the operating frequency band (62-65 GHz) for accurate RVSM up to 3 m distance. The patient under continuous health monitoring has a flexibility to change his/her position within 1 m angular distance from the fixed sensor position. Detailed RVSM measurements have been carried out with the proposed antenna system at 2 m and 3 m distances from the subject to verify the designed antenna's suitability for dynamic RVSM application. It has been shown that both the breathing rate (BR) and heartbeat rate (HR) can be successfully detected within 20° of the angular range by tuning the operating frequency from 62 to 65 GHz.
No takes yet. Share an insight, caveat, or question.
Rabbani et al. (2021) studied Remote vital sign monitoring. Fabry-Perot leaky-wave antenna (LWA) at 60 GHz-band was evaluated on Detection of breathing rate (BR) and heartbeat rate (HR). A 60 GHz-band Fabry-Perot leaky-wave antenna successfully detected breathing and heartbeat rates within a 20° angular range at 2 to 3 meters distance.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: