Key result
Leadless atrial pacemaker restores AV synchrony and eliminates atrial arrhythmia burden using existing lead sensing.
Why the study?
Integrating leadless atrial pacing with existing cardiac devices provides a potential solution for patients with atrial lead failure when lead extraction is not feasible or too risky.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
Integration of a leadless atrial pacemaker with an existing cardiac device provides a feasible and effective bailout strategy to restore atrioventricular synchrony in high-risk patients with lead failure for whom lead extraction is contraindicated.
BACKGROUND: Leadless atrial pacing provides a novel solution for patients with atrial lead failure where extraction is not feasible. The integration of leadless systems with existing cardiac devices offers a promising approach for managing patients in whom lead extraction is considered too risky. CASE SUMMARY: A 73-year-old man with heart block, after having undergone cardiac resynchronization therapy pacemaker implantation 10 years ago, presented with atrial lead failure. Owing to recent double-lung transplantation and high surgical risk, lead extraction was avoided. A leadless atrial pacemaker was successfully implanted, restoring atrioventricular synchrony by using the atrial sensing from the existing right atrial lead. DISCUSSION: This case demonstrates the integration of conventional right atrial lead sensing with leadless pacing technology, offering a safe and effective alternative when lead extraction is too risky. TAKE-HOME MESSAGE: Leadless atrial pacing integrated with existing cardiac device leads offers a safe and effective alternative for restoring atrioventricular synchrony in high-risk patients in whom lead extraction or revision is not feasible.
No takes yet. Share an insight, caveat, or question.
Goldar et al. (2025) conducted a case report in Atrial lead failure and complete heart block (n=1). Leadless atrial pacemaker (Aveir AR) was evaluated on Restoration of atrioventricular synchrony and atrial sensing. A leadless atrial pacemaker successfully restored atrioventricular synchrony and reduced atrial arrhythmia burden to zero at 6 months by utilizing an existing right atrial lead for sensing.