Answers are grounded in this paper and cite their sources.
Measurements have been made with a micro ionization chamber, of the back-scatter produced by X-ray beams of less than 10 cm. diameter in the quality range H.V.L.=0·06 mm. Cu to 1·4 mm. Cu. A table of values for back-scatter in a water phantom from these small beams is given. Values for the intermediate beam diameter of 10 cm. are compared with those previously published; and causes of discrepancy are discussed under the headings (1) Phantom material and (2) Radiation cut off by the stem of the measuring ionization chamber.
No takes yet. Share an insight, caveat, or question.
D. V. Mabbs (1949) studied this question.