Bioprosthetic valve fracture was successful in 60% of tested valves, yielding inflow area gains of 12.3% for metal stent frames and 3.6% for polymer stent frames.
Why the study?
Bioprosthetic valve fracture during valve-in-valve TAVR optimizes transcatheter heart valve expansion and reduces patient-prosthesis mismatch, but optimal balloon size and pressure require clarification.
What are the optimal balloon sizes and pressures for bioprosthetic valve fracture, and how do physical properties affect valve expansion?
What are the optimal balloon sizes and pressures for bioprosthetic valve fracture, and how do physical properties affect valve expansion?
Gains in valve area after bioprosthetic valve fracture depend on the physical properties of the surgical heart valve, with metal stent frames yielding greater expansion than polymer stent frames due to elastic recoil.
Answers are grounded in this paper and cite their sources.
Stent frame material modulates expansion after fracture in preclinical models; leaves open optimal parameters for clinical valve-in-valve procedures.
Background: Bioprosthetic valve fracture (BVF) during valve-in-valve TAVR (transcatheter aortic valve replacement) is a procedural adjunct designed to optimize the expansion of the transcatheter heart valve and reduce patient-prosthesis mismatch by using a high-pressure balloon to intentionally fracture the surgical heart valve (SHV). Methods: We performed bench testing on 15 bioprosthetic SHV to examine the optimal balloon size and pressure for BVF. We assessed morphological changes and expansion of SHV by computed tomography angiography. Successful BVF was defined as balloon waist disappearance on fluoroscopy and/or sudden pressure drop during balloon inflation. Results: Nine valves met the definition of BVF, 3 of which were confirmed by disruption of the stent frame. We classified surgical valves into 3 subsets: 1) fracturable with metal stent frame (MSF), 2) fracturable with polymer stent frame (PSF) and 3) nonfracturable. In general, valves with MSF were fractured using a balloon size = true internal diameter plus 3-5 mm inflated at high pressure (16-20 ATM) whereas valves with PSF could be fractured with a balloon size = true internal diameter plus 3-5 mm and lower balloon pressure (6-14 ATM). Gains in computed tomography angiography derived inflow area after BVF were 12.3% for MSF and 3.6% for PSF SHV. Conclusions: Gains in CT-determined valve area after BVF depend on the physical properties of the SHV, which in turn influences pressure thresholds and balloon sizing strategy for optimal BVF. Elastic recoil of PSF valves limits the gains in inflow area after BVF.
No takes yet. Share an insight, caveat, or question.
Hashimoto et al. (2024) studied Bioprosthetic valve fracture during valve-in-valve TAVR (n=15). Bioprosthetic valve fracture (BVF) was evaluated on Successful BVF (balloon waist disappearance on fluoroscopy and/or sudden pressure drop during balloon inflation). Bioprosthetic valve fracture was successful in 60% of tested valves, yielding inflow area gains of 12.3% for metal stent frames and 3.6% for polymer stent frames.