Answers are grounded in this paper and cite their sources.
Reactive oxygen species, particularly hydrogen peroxide, play a key role in cancer biology and disease progression. However, reliable tools for monitoring H₂O₂ dynamics in complex cellular models remain limited, particularly those combining analytical performance with accessibility and operational simplicity. To address this need, a low-cost paper-based analytical device (PAD) was developed for dual readouts, integrating electrochemical (ePAD) and smartphone-assisted optical/colorimetric (oPAD) formats. Both approaches employ cerium oxide nanoparticles (nCe) as nanozymes for H₂O₂ detection. With high analytical performance in both electrochemical (3 nM–500 µM linear range, LOD 0.4 nM) and optical (1–1000 µM linear range, LOD 0.3 µM) modes, covering physiologically relevant H₂O₂ concentrations, the dual-readout strategy enables independent confirmation of H₂O₂ levels in highly complex biological systems under rapid analysis (<1 min), requiring only 5 µL of sample and exhibiting excellent operational stability. Validation in 2D cultures and 3D CLL co-culture systems demonstrates the PAD capabilities to monitor redox dynamics in tumour microenvironment models. These results highlight the potential of integrating dual-readout PADs for reliable redox analysis in biologically relevant systems, paving the way for decentralized redox sensing and broader applications in biomedical research.
No takes yet. Share an insight, caveat, or question.
Gallo-Orive et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: