Answers are grounded in this paper and cite their sources.
Laser-induced breakdown spectroscopy (LIBS) in an ultra-high vacuum is used in simultaneous quantitative elemental analysis of NIST transition metal alloy samples. A plasma is formed by focusing a Nd:YAG laser onto the sample's surface inside a vacuum chamber. UV-visible emission from excited species is monitored with the use of an optical multichannel analyzer (OMA). Linear calibration curves are shown for the elements (percent composition) Al (0.2–1.2%), Cu (0.021–0.49%), Fe (4.5–51.0%), Ni (30.8–80.3%), and Zn (6–12.8%) with the use of nonresonant lines. Detection limits (signal-to-noise = 3) vary with sample composition complexity from 0.0001% for Ni in a simple copper alloy (SRM 1111) to 0.16% for Al in a complex granular sample (SRM 349a). Absolute detection limits are estimated to be in the 20–200 μg/g range for the elements of interest. Simultaneous mass spectra were taken by sweeping the magnetic field of a mass spectrometer. Preliminary results showed the magnet could not be swept fast enough for multielement analysis. The use of a position-sensitive ion detection system is proposed.
No takes yet. Share an insight, caveat, or question.
Thiem et al. (1994) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: