Key result
AMPLATZER Vascular Plug II between two MitraClips reduces intra-clip MR to mild without mitral stenosis.
Why the study?
Does implantation of an AMPLATZER Vascular Plug II between two MitraClips reduce residual intra-clip mitral regurgitation in a patient without room for a further clip?
Case Report (n=1)
Does implantation of an AMPLATZER Vascular Plug II between two MitraClips reduce residual intra-clip mitral regurgitation in a patient without room for a further clip?
This first-in-man case demonstrates the feasibility of using an AMPLATZER Vascular Plug II to treat significant residual intra-clip mitral regurgitation when an additional MitraClip cannot be placed.
Answers are grounded in this paper and cite their sources.
May offer bailout for residual intra-clip MR when further clipping is impossible; hypothesis-generating and requires prospective validation.
AIMS: We present here the first-in-man case report of a patient with a significant residual "intra-clip" jet treated by the implantation of an AMPLATZER Vascular Plug II. METHODS AND RESULTS: A 64-year-old gentleman came to our attention for mitral regurgitation (MR) recurrence four years after transcatheter mitral repair with MitraClip. After the implantation of two further MitraClips, residual intra-clip MR was present, without room for a further clip. An AMPLATZER Vascular Plug II was successfully deployed between the two clips. Final echocardiography showed residual mild MR, without any sign of mitral stenosis. CONCLUSIONS: This case demonstrates the feasibility of the implantation of an AMPLATZER Vascular Plug between two MitraClips to treat significant residual "intra-clip" MR, when the implantation of an adjunctive clip is not feasible. This approach may represent an effective therapeutic solution in case of a significant residual intra-clip jet.
No takes yet. Share an insight, caveat, or question.
Taramasso et al. (2016) conducted a case report in Residual intra-clip mitral regurgitation (n=1). AMPLATZER Vascular Plug II was evaluated on Residual mitral regurgitation and mitral stenosis on echocardiography. Implantation of an AMPLATZER Vascular Plug II between two MitraClips successfully treated significant residual intra-clip mitral regurgitation, resulting in mild residual MR without mitral stenosis.
Synapse has enriched 10 related papers. Consider them for comparative context: