Answers are grounded in this paper and cite their sources.
The purpose of this study is to develop magnetic particle imaging (MPI) technique in directly measuring time-varied cerebral superparamagnetic iron oxide nanoparticles (SPIOs) concentration in rhesus macaque. A hand-held MPI detector was developed to monitor MPI signal changes at the third harmonics of the drive frequency in resting-state nonhuman primates. Phantom experiments were firstly performed to determine the sensitivity limits of the detector as a function of distance from the detector and SPIOs concentration. Measured sensitivity profile was then used to reveal the most sensitive region of the detector. MPI detection was continuously performed to monitor MPI signal changes after two bolus injections of SPIOs in the rhesus macaque. We successfully developed a hand-held MPI to detect cerebral SPIO concentration changes in living nonhuman primate. The detection limit of the MPI detector is about 125 ng iron. We reported on the first in vivo measurement of cerebral SPIO concentration changes in rhesus macaque using a hand-held MPI detector. In vivo experiments showed the feasibility of the detector to sensitively measure MPI signals in nonhuman primate brain.
No takes yet. Share an insight, caveat, or question.
Hui et al. (2023) studied this question.