Key result
New non-Lüer connectors aim to reduce wrong-route injections in regional anesthesia versus universal Lüer connectors.
Why the study?
The universality of Lüer connectors led to serious incidents, permanent harm, and death from crossover injections via the wrong route, warranting preventative non-Lüer connector systems.
The introduction of ISO 80369-6 non-Lüer connectors aims to reduce the risk of inadvertent wrong-route injections in regional anaesthesia.
Answers are grounded in this paper and cite their sources.
May warrant adoption of ISO 80369-6 connectors to curb wrong-route errors; leaves open prospective safety trials before routine use.
Health service professionals do not have perfect vigilance. There will always be the possibility that a serious untoward incident can occur due to crossover injections of medication being administered via the wrong route. The universality of the Lüer connector made it very easy to connect several medical devices for different therapeutic uses. Although this was considered a bonus, it also showed the downside as a string of incidents, leading to permanent harm and even death, occurred, warranting actions to prevent the chance of misconnections and wrong-route injections. The International Organisation for Standardisation introduced specific ISO 80 369 connectors for use in different medical systems, with the ISO 80369-6 series of standards specifically designed to improve safety in regional anaesthesia. Human error can never be completely eliminated, but the risk of a patient inadvertently suffering from an inadvertent wrong-route injection can be reduced through the use of new 'fit and no-fit' technology. This article focuses on the universality of the Lüer connector and the introduction of the newly engineered non-Lüer connectors, for specific use in medical systems.
No takes yet. Share an insight, caveat, or question.
Wiepking et al. (2021) conducted a review in regional anaesthesia. non-Lüer connectors (ISO 80369-6) vs. Lüer connectors was evaluated. Newly engineered non-Lüer connectors (ISO 80369-6) are introduced to reduce the risk of inadvertent wrong-route injections in regional anaesthesia compared to universal Lüer connectors.
Synapse has enriched 10 related papers. Consider them for comparative context: