Key result
A transcranial magnetic stimulation device combined with a permanent magnet successfully induced shear waves of 5 and 0.5 micrometers in a cryogel phantom and chicken breast sample, respectively.
This study demonstrates the first observation of remotely induced shear waves in soft tissues using a TMS device and a permanent magnet, establishing a framework for alternative shear wave elastography.
Answers are grounded in this paper and cite their sources.
No immediate clinical role for remote shear wave induction; leaves open its translation to noninvasive elastography.
This study presents the first observation of shear waves induced remotely within soft tissues. It was performed through the combination of a transcranial magnetic stimulation device and a permanent magnet. A physical model based on Maxwell and Navier equations was developed. Experiments were performed on a cryogel phantom and a chicken breast sample. Using an ultrafast ultrasound scanner, shear waves of respective amplitudes of 5 and 0.5 μm were observed. Experimental and numerical results were in good agreement. This study constitutes the framework of an alternative shear wave elastography method.
No takes yet. Share an insight, caveat, or question.
A 2016 study studied this question. Transcranial magnetic stimulation (TMS) device combined with a permanent magnet was evaluated on Amplitude of induced shear waves. A transcranial magnetic stimulation device combined with a permanent magnet successfully induced shear waves of 5 and 0.5 micrometers in a cryogel phantom and chicken breast sample, respectively.