Answers are grounded in this paper and cite their sources.
Quantum-dot (QD) based nanosensors can be used to detect a wide range of molecules. This study examined a nanosensor comprised of thrombin binding aptamer (TBA) with 700NC InGaP QD on the 5 (') terminus and an Au nanoparticle quencher on the 3 (') terminus. Both K(+) and Pb(2+) bind to TBA, resulting in a conformational change that brings the Au quencher closer to the QD. Photoluminescence measurements indicated a decrease in fluorescence corresponding to an increase in either K(+) or Pb(2+) concentration. For healthy blood serum K(+) concentrations (3.5-5 mM), the beacon exhibited 15-17% quenching efficiency. Pb(2+) concentration of 0.48 μM, the threshold for toxicity in serum, yielded 14% quenching. The beacon's ability to detect changes in ion levels in a critical range of concentrations can make it an effective diagnostic tool.
No takes yet. Share an insight, caveat, or question.
Meshik et al. (2014) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: