Key result
Nonradioactive Rb+ efflux assay delivers highest sensitivity and lowest false-hit rate for detecting hERG inhibitors.
Answers are grounded in this paper and cite their sources.
The nonradioactive Rb+ efflux assay provides the most sensitive and accurate high-throughput method for evaluating hERG channel inhibition compared to dye-based assays.
Three functional hERG channel assay methods have been developed and evaluated. The methods were tested against five known hERG channel inhibitors: dofetilide, terfenadine (Seldane), sertindole (Serdolect), astemizole (Hismanal), and cisapride (Propulsid). The DiBAC4(3)-based assays were found to be the most economical but had high false-hit rates as a result of the interaction of dye with the test compounds. The membrane potential dye assay had fewer color-quenching problems but was expensive and still gave false hits. The nonradioactive Rb+ efflux assay was the most sensitive of all the assays evaluated and had the lowest false-hit rate.
No takes yet. Share an insight, caveat, or question.
Tang et al. (2001) studied hERG channel inhibition. Functional hERG channel assay methods was evaluated on Assay sensitivity and false-hit rate. The nonradioactive Rb+ efflux assay was the most sensitive and had the lowest false-hit rate among three evaluated methods for detecting hERG channel inhibitors.