Key result
PIABP insertion linked to ~13% vascular complication rate, mostly due to balloon removal technique.
Why the study?
What is the incidence and nature of vascular complications associated with percutaneous intra-aortic balloon pumping?
Observational (n=112)
What is the incidence and nature of vascular complications associated with percutaneous intra-aortic balloon pumping?
Percutaneous intra-aortic balloon pumping is associated with significant vascular complications, often related to the technique of balloon removal, which may require surgical intervention.
Vascular complications warrant technique-specific caution during IABP removal; leaves open the need for prospective safety studies.
A percutaneous intra-aortic balloon pump (PIABP) was inserted in 112 patients. Successful placement was achieved in 102 patients, 67 of whom survived long enough to have the device removed. Vascular complications related to the PIABP were noted in 15 patients. Reversal of ischemic signs followed PIABP removal in six patients, but nine required femoral artery exploration for thrombectomy, repair of femoral laceration, or repair of false aneurysm. No patient died as a result of PIABP or correction of associated vascular complications, though five of the patients with vascular problems died of cardiac complications. Nine of ten survivors in this group were asymptomatic, and one had persistent paresthesias six months postoperatively. Most clinically significant vascular complications were due to the technique of balloon removal, and several modifications that were effected recently are expected to decrease the incidence of complications requiring surgical intervention.
No takes yet. Share an insight, caveat, or question.
George Todd (1983) conducted an observational in Patients requiring intra-aortic balloon pumping (n=112). Percutaneous intra-aortic balloon pump (PIABP) was evaluated on Vascular complications related to PIABP. Percutaneous intra-aortic balloon pump insertion was associated with vascular complications in 13.4% (15 of 112) of patients, mostly due to the technique of balloon removal.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: