A design study was made of a neutron small-angle scattering (SAS) instrument for use at Argonne National Laboratory's proposed Intense Pulsed Neutron Source, IPNS-II (Carpenter & Werner, 1976; Werner, 1977; Carpenter, 1977). The instrument design incorporates several features: a converging-slit collimator, wavelength band-limiting choppers, capacity for large samples, and a two-dimensional detector. Machine design-dependent parameters are considered in a Monte-Carlo code that produces estimates of the instrumental resolution function and available flux at the sample. It was found that the calculated flux on sample exceeds that at steady-state reactor instruments of comparable resolution.