Variables such as scatter-air ratios and scatter functions which describe the radiation scattered to a location of interest within a patient are defined with the implicit assumption that only primary radiation impinges on the patient. Radiation scattered from the source and collimator are not considered and the dose rate in air is assumed to remain constant with changes in field size. In practice, significant scattered radiation impinges upon the patient, and dose rate in air varies with field size in a manner described by the output factor. The output factor should be incorporated into definitions of variables which describe the scattered radiation component to the total dose delivered. Output factors are incorporated into variables describing scattered radiation for both 60 Co gamma -rays and megavoltage X-rays.