Key result
Shear wave elastography yields higher variability in tissue-mimicking vessel phantoms versus water-submerged models.
Why the study?
The impact of arterial geometry, including wall thickness, pulsation, and imaging planes, on shear wave elastography estimates and measurement variability had not been thoroughly investigated.
Shear wave elastography measurements of arterial phantoms show significant variability depending on surrounding material and wall thickness, suggesting a need for combined imaging planes for accurate vessel assessment.
Requires caution interpreting vessel elastography amid tissue; leaves open standardized multi-plane protocols for clinical translation.
OBJECTIVE: Shear wave elastography imaging (SWE) is a non-invasive US technique that has been developed to provide quantitative information about tissue elasticity. This technique might be useful in the identification of vascular risk factors. Arterial wall thickness and inner diameter vary with age and disease, which may impact shear wave propagation. The effect of arterial geometry on SWE has not yet been thoroughly investigated. Therefore, this study aimed to investigate the impact of different wall thickness, pulsation and imaging planes on YM estimates, to gain more information about the source of variability associated with SWE. METHODS: Poly(vinyl alcohol) cryogel (PVA-c) fabrication has been used for phantom design and construction. The agar-based TMM was used to surround the tubes. The inlet and outlet of the phantom were connected to a programmable gear pump using c-flex tubing to form a closed loop. Image J profiling was used to clarify the anomalies further detected using SWE. RESULTS: The 4 F/T cycle vessel phantom has shown less YM variability than in the 6 F/T cycle. YM ranged from 8 kPa for a 1 mm thickness tube to 53 kPa for the thickest 6 mm wall thickness for the softer 4 F/T cycle tube. Vessel phantoms embedded in TMM show higher variability than vessel phantoms submerged in water. YM ranged from 32 kPa for a 1 mm thickness tube to 117 kPa for the thickest 6 mm wall thickness for the softer 4 F/T cycle tube. CONCLUSION: SWE variability in measurements was higher in phantoms embedded in TMM compared to those submerged in water. It is recommended that combine the transverse and longitudinal imaging planes to provide a better understanding of disease over the full vessel circumference.
No takes yet. Share an insight, caveat, or question.
Almutairi et al. (2023) studied this question. Shear wave elastography imaging (SWE) was evaluated on Young's Modulus (YM) estimates variability. Shear wave elastography variability was higher in vessel phantoms embedded in tissue-mimicking material (YM 32-117 kPa) compared to those submerged in water (YM 8-53 kPa).