Key result
Active-fixation ICD lead perforates septum and LV requiring emergency surgical extraction in HCM patient.
Why the study?
Left ventricular perforation caused by an ICD lead is a rare but potentially fatal complication.
Case Report (n=1)
Hypertrophic myocardium does not preclude life-threatening ICD lead perforation, underscoring the importance of early recognition and individualized surgical management.
Answers are grounded in this paper and cite their sources.
LV perforation may occur days after septal ICD placement in HCM; single case leaves open incidence and prevention strategies.
Left ventricular (LV) perforation caused by an implantable cardioverter-defibrillator (ICD) lead is a rare but potentially fatal complication. We describe the case of a 34-year-old man with hypertrophic cardiomyopathy who underwent single-chamber ICD implantation for primary prevention of sudden cardiac death using an active-fixation transvenous lead positioned in the right ventricular septum. Four days after implantation, he developed acute chest pain, dyspnoea, hypotension, with clinical signs of evolving cardiac tamponade. Transthoracic echocardiography demonstrated a moderate circumferential pericardial effusion, while device interrogation revealed abnormal pacing parameters, prompting emergency pericardiocentesis with immediate haemodynamic stabilization. Computed tomography was subsequently performed to accurately delineate lead position and confirmed perforation through the interventricular septum with extension into the LV free wall, a finding that could not be fully characterized by echocardiography alone. The patient was transferred to a tertiary centre with continuous cardiac surgical support, where lead extraction and surgical repair of the myocardial defects were successfully performed using pledgeted sutures. The postoperative course was uneventful and follow-up imaging confirmed complete resolution of the pericardial effusion. Unlike the typical pattern of right ventricular free-wall perforation, this case involved transseptal migration of an ICD lead with left ventricular involvement, highlighting that hypertrophic myocardium does not preclude this life-threatening complication and underscoring the importance of early recognition and individualized management.
No takes yet. Share an insight, caveat, or question.
Alexiou et al. (2026) conducted a case report in Hypertrophic cardiomyopathy with ICD lead perforation (n=1). ICD implantation with active-fixation transvenous lead was evaluated. An active-fixation ICD lead caused a rare transseptal and left ventricular perforation in a 34-year-old man with hypertrophic cardiomyopathy, necessitating emergency pericardiocentesis and surgical extraction.
Synapse has enriched 10 related papers. Consider them for comparative context: