PURPOSE: To evaluate magnetic field interactions for 109 different biomedical implants and devices in association with exposure to a 3.0-Tesla magnetic resonance (MR) system. MATERIALS AND METHODS: A total of 109 implants and devices (aneurysm clips, 32; clips, fasteners, and staples, 10; coils and stents, 10; heart valve prostheses and annuloplasty rings, 12; orthopedic implants, five; suture materials, 13; vascular access ports and accessories, 13; miscellaneous implants and devices, 14) were tested for magnetic field interactions at 3.0-Tesla using previously-described, standardized techniques to assess magnetic field translational attraction and torque. RESULTS: The deflection angles and torque measurements ranged, respectively, from 0 to 16 degrees and 0 to +2 for the aneurysm clips; 0 to 90 degrees and 0 to +4 for the clips, fasteners, and staples; 0 to 47 degrees and 0 to +4 for the coils and stents; 0 to 4 degrees and 0 to +1 for the heart valve prostheses and annuloplasty rings; 0 to 12 degrees and 0 to +2 for the orthopedic implants; 0 to 13 degrees and 0 to +2 for the suture materials; 0 to 52 degrees and 0 to +4 for the vascular access ports and accessories; and 0 to 28 degrees and 0 to +3 for the miscellaneous implants and devices. CONCLUSION: Of the 109 implants and devices assessed for magnetic field interactions at 3.0-Tesla, four (4%) are potentially unsafe based on deflection angle criteria. The implications of these results for patients undergoing MR procedures at 3.0-Tesla is discussed. Notably, these results are specific to the 3.0-Tesla MR system used for this evaluation.
No takes yet. Share an insight, caveat, or question.
Frank G. Shellock (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: