Key result
Stepwise implantation achieves 96% His bundle pacing success in patients with right atriomegaly.
Why the study?
An enlarged right atrium poses an anatomical challenge for His bundle pacing, and whether new implantation tools can overcome these challenges was unknown.
Does the availability of different dedicated delivery sheaths improve procedural success of permanent His bundle pacing in patients with right atriomegaly?
Observational (n=24)
Does the availability of different dedicated delivery sheaths improve procedural success of permanent His bundle pacing in patients with right atriomegaly?
The availability of multiple dedicated delivery sheath curves and lead types improves the procedural success of His bundle pacing in patients with challenging anatomy such as right atriomegaly.
May aid His bundle pacing in right atriomegaly; single case report leaves open prospective validation.
BACKGROUND: An enlarged right atrium (RA) is a challenging anatomy that can limit the successful use of His bundle pacing (HBP). It is unknown whether new implantation tools could help overcome these challenges. METHODS: in women underwent permanent HBP. We used a stylet-driven lead (SDL) with an extendable helix delivered via a dedicated delivery sheath (Selectra 3D, Biotronik) as a first attempt. In case of failure, a second attempt was performed with the same lead but with a different delivery curve. Finally, a lumen-less lead (LLL) was also available as a third attempt. RESULTS: (45-54). Using SDL, HBP was achieved with a single sheath curve in 17 patients (71%). The second attempt with the same lead but a different sheath was successful in four more patients (SDL success 87%). The fluoroscopy time increased significantly when the second attempt was necessary (8 min [6-11] vs. 15 min [13-17], p < .001). In the remaining three patients, HBP was further attempted with a LLL leading to a final procedural success of 96%. No lead dislodgment nor significant increase in pacing threshold was observed at 1-month (1.2 [0.7-1.7] V@1.0 ms vs. 1.1 [0.8-1.7] V@1.0 ms, p = .939). CONCLUSION: The availability of different dedicated delivery systems for HBP can improve procedural outcomes even in challenging circumstances, such as in patients with right atriomegaly.
No takes yet. Share an insight, caveat, or question.
Marinaccio et al. (2023) conducted an observational in Right atriomegaly (n=24). Different dedicated delivery sheaths and leads for His bundle pacing was evaluated on Final procedural success of His bundle pacing. A stepwise implantation approach using different dedicated delivery sheaths and leads achieved a final His bundle pacing procedural success rate of 96% in patients with right atriomegaly.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: