Key result
STAN monitoring shows severe limitations detecting hypoxemia, losing ST-waveforms during ~62% of umbilical cord occlusion.
Why the study?
Does the STAN device accurately detect hypoxemia in a fetal sheep model exposed to prolonged umbilical cord occlusion?
Does the STAN device accurately detect hypoxemia in a fetal sheep model exposed to prolonged umbilical cord occlusion?
Absolute Event Rate: 62% vs 14%
p-value: p=<0.05
The STAN device demonstrated significant signal processing limitations, failing to assess ST-waveforms during 62% of prolonged umbilical cord occlusion in a fetal sheep model.
STAN limitations in fetal sheep hypoxemia model caution clinical reliance; leaves open need for device refinements and human validation.
INTRODUCTION: The inconclusive clinical results for ST-waveform analysis (STAN) in detecting fetal hypoxemia may be caused by the signal processing of the STAN-device itself. We assessed the performance of a clinical STAN device in signal processing and in detecting hypoxemia in a fetal sheep model exposed to prolonged umbilical cord occlusion (UCO). METHODS: Eight fetal lambs were exposed to 25 minutes of UCO. ECG recordings were analyzed during a baseline period and during UCO. STAN-event rates and timing of episodic T/QRS rise, baseline T/QRS rise and the occurrence of biphasic ST-waveforms, as well as signal loss, were assessed. RESULTS: During baseline conditions of normoxemia, a median of 40 (IQR, 25-70) STAN-events per minute were detected, compared to 10 (IQR, 2-22) during UCO. During UCO STAN-events were detected in five subjects within 10 minutes and in six subjects after 18 minutes, respectively. Two subjects did not generate any STAN-event during UCO. Biphasic ST event rate was reduced during UCO (median 0, IQR 0-5), compared to baseline (median 32, IQR, 6-55). ST-waveforms could not be assessed in 62% of the recording time during UCO, despite a good quality of the ECG signal. CONCLUSIONS: The STAN device showed limitations in detecting hypoxemia in fetal sheep after prolonged UCO. The STAN device produced high false positive event rates during baseline and did not detect T/QRS changes adequately after prolonged fetal hypoxemia. During 14% of baseline and 62% of the UCO period, the STAN-device could not process the ECG signal, despite its good quality. Resolving these issues may improve the clinical performance of the STAN device.
No takes yet. Share an insight, caveat, or question.
Andriessen et al. (2018) studied Fetal hypoxic distress (n=8). Prolonged umbilical cord occlusion vs. Baseline (normoxemia) was evaluated on ST waveform signal loss (% of recording time) (p=<0.05). The STAN device showed severe limitations in detecting hypoxemia, with ST-waveforms unassessable in 62% of the recording time during umbilical cord occlusion compared to 14% during baseline.