Key result
Inferior interatrial septum lead insertion with a preshaped stylet achieves 95% success without complications.
Why the study?
Does a technique using a preshaped stylet and fluoroscopic guidance allow safe and successful placement of a pacing lead at the inferior portion of the interatrial septum in patients with pacing indications and atrial fibrillation?
Observational (n=117)
Does a technique using a preshaped stylet and fluoroscopic guidance allow safe and successful placement of a pacing lead at the inferior portion of the interatrial septum in patients with pacing indications and atrial fibrillation?
Placement of an active fixation pacing lead at the inferior portion of the interatrial septum using a preshaped stylet is highly successful and safe without requiring special tools.
Answers are grounded in this paper and cite their sources.
May support preshaped stylet use for inferior interatrial septal pacing in AF; leaves open randomized outcome trials before practice change.
INTRODUCTION: Previous studies have suggested that, among septal sites, the inferior portion of the interatrial septum (IAS) is the most likely to prevent atrial fibrillation, though inserting an active fixation lead at this site can be tedious and time consuming. We describe a relatively straightforward technique to insert a lead at this site without special tools. METHOD: We studied 117 consecutive patients (mean age = 76 +/- 8 years, 69% men) with ACC/AHA class I and II pacing indications and histories of paroxysmal or permanent atrial fibrillation, undergoing implantation of a dual chamber pacing system. A technique using the "preshaped" stylet and fluoroscopic guidance is described. RESULTS: The insertion was successful in 111 patients (95%). Acute dislodgement occurred in six patients (5%). The intrinsic P-wave duration was 117 +/- 22 ms, and the paced P-wave duration was 90 +/- 20 ms (23% shortening, P < 0.001). The mean time required to insert the atrial lead was 12 +/- 8 minutes. No complications occurred. CONCLUSIONS: Insertion of an active fixation lead at the inferior portion of the interatrial septum was safe and highly successful in the majority of patients with this technique.
No takes yet. Share an insight, caveat, or question.
Acosta et al. (2007) conducted an observational in ACC/AHA class I and II pacing indications and atrial fibrillation (n=117). Active fixation lead insertion at the inferior portion of the interatrial septum using a preshaped stylet was evaluated on Successful insertion. Insertion of an active fixation lead at the inferior portion of the interatrial septum using a preshaped stylet was successful in 95% of patients, with no complications.
Synapse has enriched 10 related papers. Consider them for comparative context: