Key result
Sutureless aortic ring connector cuts anastomosis time by ~39% versus manual suturing in swine.
Why the study?
Does a novel ring connector for sutureless aortic anastomosis reduce anastomosis time and maintain patency compared to traditional manual suture in a swine model?
Does a novel ring connector for sutureless aortic anastomosis reduce anastomosis time and maintain patency compared to traditional manual suture in a swine model?
Absolute Event Rate: 9.47% vs 15.58%
A novel ring connector for sutureless aortic anastomosis demonstrates feasibility, reducing anastomosis time while maintaining patency in a preclinical swine model.
Reduces aortic anastomosis time in swine; hypothesis-generating and requires human trials before any clinical consideration.
The surgical treatment for aortic diseases remains a challenge for any cardiac surgeon. The use of sutureless ring connector in aortic anastomosis can simplify the procedure and shorten anastomosis time. Therefore, we developed a novel device for sutureless aortic anastomosis. A series of experiments were carried out for tensile and leakproof-capacity assessments to verify the feasibility of the ring connector by using fresh swine aorta samples. In in vivo test, the ring connector was implanted in 6 swine with follow-up of 6 months. Radiographic and pathological studies of the aorta were performed. In the tensile tests, the strength was 32.7±5.9 Newton (N) in the sutureless anastomosis group, compared with 73.3±12.5 N in the control group by traditional manual suture. In the leakproof-capacity assessment, no sign of either leakage or bursting was evident at 280 mmHg of internal pressure in the aorta samples. In in vivo tests, it took 9.47±0.3 minutes for the sutureless anastomosis, compared with 15.58±1.39 minutes for hand-sewn suturing. Insertion was easy and rapid. Radiographic and pathological studies were performed at first month, third month and sixth month after surgery, each time obtained from the two swine, showed patency of the anastomosis and no signs of stenosis, blood leakage, migration or pseudoaneurysm formation, except one paralyzed swine developed of thrombo-occlusion at the site of the sutureless anastomosis. The result indicates that this novel ring connector offers considerable promise for sutureless aortic anastomosis.
No takes yet. Share an insight, caveat, or question.
Shao et al. (2017) studied Aortic anastomosis (n=6). Novel ring connector for sutureless aortic anastomosis vs. Traditional manual suture was evaluated on Anastomosis time (minutes). Sutureless aortic anastomosis using a novel ring connector required 9.47±0.3 minutes compared with 15.58±1.39 minutes for traditional hand-sewn suturing in a swine model.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: