Key result
Suspended weight device proposed to correct Caesarean table tilt averaging only ~8° in practice.
A simple suspended weight device can help accurately set a 15° left lateral tilt during Caesarean sections, addressing the common underestimation of tilt angles by staff.
Answers are grounded in this paper and cite their sources.
May compromise aortocaval decompression in Caesarean sections; leaves open whether a suspended weight device improves positioning accuracy or outcomes.
We read with interest the recent article (Mendonca et al. Anaesthesia 2003; 58: 428–31) that reported the effect of different degrees of left lateral tilt to prevent hypotension following regional anaesthesia for Caesarean section. We were especially interested in their discussion of the amount of tilt necessary to prevent or treat aortocaval compression. We recently conducted an audit in our department of actual and estimated table tilt angles during elective Caesarean section. Measured angles (mean actual angle = 8.2°, SD 3°) were well below the recommended 15°, although 73% of audited staff (n = 15) quoted 15° correctly as being the most commonly recommended angle. Our obstetric anaesthetists also generally overestimated the angle they had set the table (mean overestimate = 5.3°, SD 3.3°). Jones et al. recently published similar results [1]. Operating tables do not usually have inbuilt inclinometers, and spirit levels measuring angles accurately are expensive. Our audit highlights the difficulty of estimating and setting consistently adequate tilt angles without any visual or mechanical reference. For the above reasons we suggest a simple and cheap device, which allows the exact left lateral tilt setting of the operating table to 15° without the need for repeated measurements or calculations. This involves a weight suspended from the left edge of the table that just reaches the ground at 15° left lateral tilt (Fig. 6). To ensure reproducible angle setting, the table must be in the lowest position, after which the table height can be adjusted according to the surgeon's preferences. Correct setting of the angle is independent of head down/up tilt, as the weight is suspended at the same level as the transverse axis of the table and therefore does not interfere with, for example, Trendelenburg positioning to influence spread of spinal anaesthesia. Initial calibration at 15° is established by calculating the necessary difference in height of the left and right table edges, which equals the sine of 15° times the table width [height right − height left = sin (15°) × table width]. The weight is suspended at the level of the transverse table axis. Some controversy exists about the degree of table tilt necessary to prevent or treat aortocaval compression. The 15° recommendation appears somewhat arbitrary, especially as it does not take patient factors such as weight or fetal size into account. Furthermore, even at 15° tilt, aortic compression has been demonstrated [2]. The degree of left lateral tilt selected will be a compromise between avoiding aortocaval compression, patient comfort and ease of surgical access. However, we feel this device is valuable as a reminder and guide for lateral tilt and is cheaper and easier to install than a spirit level or inclinometer.
No takes yet. Share an insight, caveat, or question.
Claas Siegmueller (2003) conducted a letter in Caesarean section (n=15). Simple device (suspended weight) for 15° left lateral tilt vs. Standard practice (no visual or mechanical reference) was evaluated on Actual table tilt angle. An audit of 15 staff members showed that actual table tilt angles during Caesarean section averaged 8.2°, well below the recommended 15°, prompting the proposal of a simple suspended weight device.
Synapse has enriched 10 related papers. Consider them for comparative context: