Answers are grounded in this paper and cite their sources.
Preclinical study demonstrates wide-range tissue modulus sensing using a shape-memory alloy probe in biological tissues, suggesting improved diagnostic capability for tumor localization.
Elastic modulus is a valuable pathological biomarker to make diagnoses. As normal tissue becomes afflicted with diseases, the local modulus changes. Here, we present a modulus characterization probe based on a shape-memory alloy (SMA) microwire integrating both sensing and actuation to quantify modulus variations. Under controlled current input, SMA microwire contracts to generate active sensing force. The extent of contraction dynamically changes in response to the tissue modulus, which is quantified through real-time resistance changes. The maximum modulus measurement range was in the GPa range, improved by three orders of magnitude compared to conventional methods. It provided a spatial resolution of 2 mm with a data resolution of 1.53 mN in the kPa range and a spatial resolution of 1 mm with a data resolution of 1.50 mN in the MPa range. The device was proved to be able to be used to target tumors and distinguish tissues in in vivo and ex vivo tests.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2025) studied this question.