Answers are grounded in this paper and cite their sources.
In recent years, a major research goal of companies has been to create mechanical components suitable for rehabilitation that are safer and more reliable. Evaluating their biomechanical features could be a way to improve them. The purpose of this study was to evaluate the different biomechanical features of low-profile retentive systems (Rhein®). Two different attachment systems were tested: OT Equator® Smart Box and Locator® R-TX. Once a machine was created for the simulation of the connection and disconnection of the attacks in a combined manner, it was possible to evaluate these parameters over time. Attachments were mounted in two different configurations of the divergence angle: 10° and 50°. The drop retention force proved to be stable over time. The Locator® R-TX attachment experienced a more rapid decrement of the retention force than the OT Equator® Smart Box. Both tested systems experienced a high drop in retention; this drop tended to stabilize after 1.5 years of use, and it was correlated with the divergence angle. The OT Equator® Smart Box system underwent this loss of retention more gradually than the Locator® R-TX. This study demonstrates preliminary results from a bioengineering and biomechanical point of view, providing useful information for the continuous improvement of these devices and, therefore, for the quality of patients’ oral health.
No takes yet. Share an insight, caveat, or question.
Cervino et al. (2019) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: