Answers are grounded in this paper and cite their sources.
AIM: Midazolam is a commonly used marker substrate for the in vivo assessment of CYP3A activity. Reliable pharmacokinetic assessment at sub-pharmacological doses of midazolam requires an ultra-sensitive analytical method. METHODS: A new, ultra-sensitive LC-MS/MS method for the determination of midazolam in human plasma using SPE was developed and fully validated. The lowest limit of quantitation is 0.1 pg/ml with a sample volume of 500 μl. RESULTS/CONCLUSION: The following parameters were validated: sensitivity, assay accuracy and precision, linearity, selectivity, and stability of midazolam at pertinent analytical and storage conditions. The validated method was utilized successfully for the sample assay during a midazolam microdosing study for the evaluation of CYP3A4 activity of a clinical candidate.
No takes yet. Share an insight, caveat, or question.
Chen et al. (2017) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: