Key result
Continuous contrast-enhanced CDU identifies ~5-fold more endoleaks than computed tomography.
Why the study?
Does continuous ultrasound contrast infusion as an adjunct to color duplex ultrasound improve the detection of endoleaks in patients undergoing aortic endograft surveillance compared to CTA?
Observational (n=18)
Does continuous ultrasound contrast infusion as an adjunct to color duplex ultrasound improve the detection of endoleaks in patients undergoing aortic endograft surveillance compared to CTA?
Continuous ultrasound contrast infusion as an adjunct to color duplex ultrasound is a reliable modality for endoleak detection after endovascular aortic aneurysm repair, potentially outperforming conventional CTA.
Answers are grounded in this paper and cite their sources.
Contrast-enhanced duplex may detect more endoleaks than CTA post-EVAR; leaves open whether it should replace CTA pending prospective trials.
Objective —We sought to evaluate continuous ultrasound contrast infusion Optison® as an adjunct to color duplex ultrasound (CDU) in endoleak detection. Methods —Endoleak surveillance, including CDU, Optison®, and computed tomography (CTA), was performed during 18 examinations. One (3 ml) vial of Optison® was diluted in a 57-ml syringe of normal saline (total 60 ml) and administered by infusion pump at 4 ml/hr. Transverse and sagittal views were performed of the aortic aneurysm sac, endograft, and arteries outside the aneurysm sac. Primary endpoints were endograft patency, the presence or absence of endoleak, and limb dysfunction. Results —There were no adverse events related to the contrast agent. Contrast appearance through the endograft occurred after an average of 1 min using a mean of 46.8 ml of contrast infusion per patient. There were 10 enodleaks clearly identified by CDU and Optison®, i.e., 2 type I, 4 type II, and 2 type III endoleaks with an additional study suspicious for endoleak. Using CDU and CTA, we found 9 studies without endoleaks whereas with Optison®, we found 8 studies normal. Comparatively, when using CTA, we identified only 2 definitive endoleaks. Of the 9 endoleaks identified by using CDU and Optison®, 2 of the endoleaks were type I and not observed on conventional CTA and were later confirmed by arteriogram. One study was deemed negative by CDU and Optison® whereas CTA results were suspicious. Body habitus was prohibitive for definitive CDU findings in one patient. Conclusions —Using continuous ultrasound contrast infusion as an adjunct to CDU allowed for longer imaging time and more extensive evaluation of the endograft, the aneurysm sac, and the perianeurysmal arteries in these patients. Our early results demonstrate that the use of ultrasound contrast for endoleak detection is reliable and may be considered a primary surveillance modality after endovascular aortic aneurysm repair.
No takes yet. Share an insight, caveat, or question.
Hodge et al. (2007) conducted an observational in Endovascular aortic aneurysm repair (n=18). Continuous ultrasound contrast infusion (Optison) with color duplex ultrasound vs. Computed tomography (CTA) was evaluated on Endograft patency, the presence or absence of endoleak, and limb dysfunction. Continuous ultrasound contrast infusion with color duplex ultrasound identified endoleaks in 55.6% of examinations, compared to 11.1% with computed tomography.