Answers are grounded in this paper and cite their sources.
The response times of the Aanderaa 4330, Aanderaa 4330 WTW, RBRcoda T.ODO|slow, and PyroScience PICO‐O2‐SUB were evaluated in the laboratory over a range of profiling speeds at two temperatures. The PyroScience PICO‐O2‐SUB had the fastest response time (1–4 s), followed by the RBRcoda T.ODO|slow (~ 15–35 s), Aanderaa 4330 (~ 30–60 s), and Aanderaa 4330W (~ 50–100 s). This study provides recommendations on improving the quality of oxygen data from optodes in profiling applications by additionally assessing the impact of response time testing setups, thermal inertia effects, and foil types on sensor response times. This study provides a new response time function based on physical principles to predict response time for these four optode types.
No takes yet. Share an insight, caveat, or question.
Park et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: