Key result
Serial 1.5T CMR scans in ICD patients show 0 complications or device dysfunctions.
Why the study?
Does serial cardiac MRI at 1.5 T cause adverse effects or device dysfunction in patients with implantable cardioverter defibrillators?
Population
10 patients with implantable cardioverter defibrillators (ICDs), median age 56 years, 9 men.
Design
Case_series
Follow-up
median 370 days (range 274-723)
Authors
Loading...
Serial MR may be feasible in ICD patients; leaves open whether guidelines should shift without randomized confirmation.
Observational (n=10)
Does serial cardiac MRI at 1.5 T cause adverse effects or device dysfunction in patients with implantable cardioverter defibrillators?
Serial 1.5 T cardiac MRI can be safely performed in patients with ICDs without adverse effects on the patient or device function, despite historical contraindications.
Junttila et al. (2011) conducted an observational in Implantable cardioverter defibrillators (ICDs) (n=10). Serial cardiac MR scans at 1.5 T was evaluated on Complications and ICD dysfunction. Serial cardiac MR scans at 1.5 T in patients with ICDs resulted in 0 complications and no occurrences of ICD dysfunction during a median follow-up of 370 days.
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: