Key result
Serial 1.5T CMR scans in ICD patients show 0 complications or device dysfunctions.
Why the study?
Does serial cardiac MRI at 1.5 T cause adverse effects or device dysfunction in patients with implantable cardioverter defibrillators?
Observational (n=10)
Does serial cardiac MRI at 1.5 T cause adverse effects or device dysfunction in patients with implantable cardioverter defibrillators?
Serial 1.5 T cardiac MRI can be safely performed in patients with ICDs without adverse effects on the patient or device function, despite historical contraindications.
Answers are grounded in this paper and cite their sources.
Serial MR may be feasible in ICD patients; leaves open whether guidelines should shift without randomized confirmation.
OBJECTIVE: While patients with cardiac implantable electronic devices could benefit from magnetic resonance (MR) imaging, the presence of such devices has been designated as an absolute contraindication to MR. Although scanning algorithms are proposed for cardiac implantable electronic devices, their safety remains uncertain. To address this issue, the safety of serial cardiac MR scans was evaluated in patients with implantable cardioverter defibrillators (ICDs). METHODS: Three serial cardiac MR scans were prospectively performed at 1.5 T on 10 patients (9 men) of median age 56 years (range 51-68) with ICDs. ICD interrogation was performed before and after the MR scan and at a follow-up of median 370 days (range 274-723). Image quality was also assessed. RESULTS: In all patients MR scanning occurred without complications. There were no differences between pre- and post-MR pacing capture threshold, pacing lead or high voltage lead impedance, or battery voltage values. During follow-up there were no occurrences of ICD dysfunction. Although most patients had image artifacts, the studies were generally diagnostic regarding left ventricular function and wall motion. Delayed enhancement imaging was of good quality for inferior wall and inferolateral infarcts, but ICD artifacts often affected the imaging of anterior wall infarcts. CONCLUSION: Serial MR scans at 1.5 T in patients with ICDs, when carefully performed in a monitored setting, have no adverse effects on either patient or device. When required, single or multiple MR scans at 1.5 T may therefore be considered for clinical diagnostic purposes in these patients.
No takes yet. Share an insight, caveat, or question.
Junttila et al. (2011) conducted an observational in Implantable cardioverter defibrillators (ICDs) (n=10). Serial cardiac MR scans at 1.5 T was evaluated on Complications and ICD dysfunction. Serial cardiac MR scans at 1.5 T in patients with ICDs resulted in 0 complications and no occurrences of ICD dysfunction during a median follow-up of 370 days.