Answers are grounded in this paper and cite their sources.
A high-performance liquid chromatographic method was developed for the simultaneous determination of 2 ternary mixtures containing amiloride hydrochloride, atenolol, hydrochlorothiazide, and chlorthalidone used in hypertension therapy. The use of cyanopropyl column results in satisfactory separation of both mixtures. The mobile phase consisted of 10 mM KH2PO4 buffer (pH 4.5) and methanol in a ratio of (75 + 25% v/v), at a flow rate of 1 mL/min. UV detector was operated at 275 nm. Calibration graphs were linear in the concentration ranges of 2-10, 20-200, 10-100, and 5-50 microg/mL for amiloride hydrochloride, atenolol, hydrochlorothiazide, and chlorthalidone, respectively. Intraday and interday precision values (relative standard deviation) were <1.13 for mixture I (amiloride hydrochloride, atenolol, chlorthalidone), and <0.93 for mixture II (amiloride hydrochloride, atenolol, hydrochlorothiazide). The method was successfully applied for the determination of the 2 combinations in laboratory-prepared mixtures and commercial pharmaceutical formulation with high accuracy and precision. Statistical comparison of the results with those of the published methods showed excellent agreement and indicates no significant difference between them.
No takes yet. Share an insight, caveat, or question.
Elshanawane et al. (2009) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: