Key result
Review links 26 cases of PTFE neo-chordae rupture after mitral valve repair to suture size.
Why the study?
The incidence of neo-chordae failure causing recurrent mitral regurgitation following mitral valve repair with expanded polytetrafluoroethylene sutures is not well defined.
What are the complications and causes of failure associated with artificial neo-chordae in mitral valve repair?
What are the complications and causes of failure associated with artificial neo-chordae in mitral valve repair?
Answers are grounded in this paper and cite their sources.
While artificial neo-chordae for mitral valve repair are generally durable, late chordal rupture can occur and is mainly related to suture size, highlighting the need for ongoing follow-up.
BACKGROUND: Mitral valve repair using expanded polytetrafluoroethylene sutures to replace mitral chordae tendineae is a well-established procedure. However, the incidence of neo-chordae failure causing recurrent mitral regurgitation is not well defined. METHODS: We have reviewed the reported cases of complications after mitral valve repair related to the use of neo-chordae. This study was mainly carried out through PubMed, Medline, and Google Chrome websites. RESULTS: We have identified a total of 26 patients presenting with rupture of polytetrafluoroethylene neo-chordae, mostly being described as isolated cases. Few other cases of recurrent mitral regurgitation with hemolysis were found, where reoperation was not caused by neo-chordal failure but most likely by technical errors. At pathological investigation the findings were substantially similar in all reported cases. The neo-chordae retained their length and pliability, became covered with host tissue and rupture was mainly related to suture size. Mild calcification was observed not interfering with chordal function; chordal infection did never occur. CONCLUSIONS: The use of artificial neo-chordae provides excellent late results with durable mitral valve repair stability. Chordal rupture may occur late postoperatively leading to reoperation because of recurrent mitral regurgitation. Despite its rarity, this potential complication should not be overlooked during follow-up of patients after mitral valve repair using artificial neo-chordae.
No takes yet. Share an insight, caveat, or question.
Vendramin et al. (2022) conducted a review in Mitral valve repair complications (n=26). Artificial neo-chordae (expanded polytetrafluoroethylene sutures) was evaluated on Rupture of polytetrafluoroethylene neo-chordae. A review identified 26 reported cases of polytetrafluoroethylene neo-chordae rupture after mitral valve repair, with pathological findings showing rupture was mainly related to suture size.
Synapse has enriched 10 related papers. Consider them for comparative context: