Answers are grounded in this paper and cite their sources.
In this paper, we report a scheme providing precise spectral analysis and surface imaging, simultaneously, based on a high-coherence dual-comb interferometer. With two tightly phase-locking frequency combs, we demonstrate a high-coherence dual-comb interferometer (DCI) covering 188 to 195 THz (1538.5 to 1595.7 nm) with comb-tooth resolution and a max spectral signal-to-noise ratio (SNR) of 159.7. The combination of the high-coherence dual-comb spectrometer and a reference arm simultaneously enables gas absorption spectroscopy and for the absolute distance information to be obtained in one measurement. As a demonstration, we measure the spectrum of CO2 and CO. From the same interferograms, we demonstrate that distance measurement, by time-of-flight (TOF), can be resolved with an rms precision of 0.53 μm after averaging 140 images and a measurement time of 1 s. Finally, we demonstrate that non-contact surface imaging, using 2D mechanical scanning, reaches lateral resolution of 40 μm. The longitudinal precision is 0.68 μm with a measurement time of 0.5 s. It verifies that DCS has the potential to be applied in standoff detection, environmental pollution monitors, and remote sensing.
No takes yet. Share an insight, caveat, or question.
Deng et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: