Key result
Measuring aortic diameter at the aortic valve yielded the most accurate left ventricular output quantification compared to MRI (bias -10.6 ml/kg/min; LOA -74 to 52).
Why the study?
The reproducibility and accuracy of left ventricular output quantifications in neonates using left ventricular outflow tract diameter measured at three anatomical landmarks were not well established.
Does measuring LVOTD at the aortic valve improve the reproducibility and accuracy of LVO calculation in neonates compared to other anatomical landmarks?
Observational (n=58)
Does measuring LVOTD at the aortic valve improve the reproducibility and accuracy of LVO calculation in neonates compared to other anatomical landmarks?
Measuring aortic diameter at the hinges of the aortic valve provides the most reproducible and accurate echocardiographic calculation of left ventricular output in neonates.
OBJECTIVES: To assess reproducibility and accuracy of left ventricular output (LVO) quantifications in neonates, when left ventricular outflow tract diameter (LVOTD) was measured at the hinges of the aortic valve (AV), at the aortic sinus (AS), and at the sinotubular junction (STJ). METHODS: This was an observational study. In the first cohort of very preterm neonates, we assessed intraobserver and interobserver repeatability of LVOTD measured at the AV, AS, and STJ and of the corresponding LVO. In the second cohort of older neonates, we compared paired LVO measurements by echo and magnetic resonance imaging (MRI). RESULTS: In the first cohort of 48 neonates, mean (standard deviation) weight and age at scan were 1046 (302) g and 28.1 (2.7) weeks. Interobserver bias (95% limits of agreement [LOA]) for LVOTD at the AV, AS, and STJ was 0 (-0.3 to 0.3) mm, 0 (-0.7 to 0.7) mm, and 0 (-0.8 to 0.7) mm, respectively. Interobserver bias (95% LOA) for the corresponding LVO was -1.3 (-31 to 33) ml/kg/min, -0.5 (-88 to 87) ml/kg/min, and -7.2 (-83 to 69) ml/kg/min, respectively. In the second cohort of 10 neonates, median (range) weight and age at scan were 1942 (970-3640) g and 37.2 (31.7-39.8) weeks. LVO measured at the AV showed stronger agreement with MRI: bias (LOA) -10.6 (-74 to 52) ml/kg/min, compared to LVO measured at AS and STJ: 194 (-0.5 to 388) ml/kg/min and 43 (-72 to 159) ml/kg/min respectively. CONCLUSIONS: Reproducibility and accuracy of LVO quantification by echo were better when aortic diameter was measured at AV.
No takes yet. Share an insight, caveat, or question.
Ficial et al. (2021) conducted an observational in Neonates requiring left ventricular output quantification (n=58). Aortic diameter measurement at the aortic valve (AV) vs. Measurement at the aortic sinus (AS) or sinotubular junction (STJ) was evaluated on Agreement of left ventricular output (LVO) measured by echo with MRI (bias -10.6 ml/kg/min, 95% CI -74 to 52). Measuring aortic diameter at the aortic valve yielded the most accurate left ventricular output quantification compared to MRI (bias -10.6 ml/kg/min; LOA -74 to 52).