Cohort study reveals anatomical and alignment factors altering J-Taper stem version in primary hip arthroplasty, indicating targets to optimize implant positioning.
Background: The appropriate combination of acetabular component and stem anteversion is needed to reduce implant impingement and dislocation of total hip arthroplasty (THA). J-Taper is a tapered wedge stem with a relatively shorter small area of arc spray coating compared with conventionally tapered wedge stems. We examined the factors that affect postoperative version of J-Taper. Methods: A total of 59 patients who underwent primary THA were assessed by computed tomography. Preoperative femoral neck and postoperative stem anteversion, stem sagittal alignment, and stem coronal alignment were measured. Results: Mean femoral neck anteversion was 26.1° ± 12.2°. Mean stem version was 32.3° ± 9.0°. Multiple linear regression analysis showed that femoral neck anteversion (p<0.01), stem sagittal alignment (p=0.045), and body height (p=0.048) were positively correlated with stem aversion. The stem version was significantly increased compared with that of native femoral anteversion (6.3° ± 10.9°). Multiple linear regression analysis showed that femoral neck anteversion was negatively correlated with an increased angle of stem version, body height, and stem sagittal alignment was positively correlated. Conclusions: Our findings are helpful for surgeons during preoperative planning to achieve the appropriate combined anteversion.
No takes yet. Share an insight, caveat, or question.
Kawakami et al. (2017) studied this question.