Answers are grounded in this paper and cite their sources.
The sensitivities of five transistorized nuclear magnetic resonance spectrometers have been measured: a Q meter, a Robinson oscillator, a common marginal oscillator, a modified PKW oscillator and a marginal oscillator the limiter of which was placed behind an r.f. amplifier. All the spectrometers except the common marginal oscillator appeared to possess the same sensitivities for all considered oscillation levels and modulation frequencies, provided that the PKW oscillator and the second marginal oscillator are made sufficiently critical. The common marginal oscillator was less sensitive as this spectrometer could not be made very critical, which resulted in a large contribution of the 1/ f noise.
No takes yet. Share an insight, caveat, or question.
Robert A. Wind (1970) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: