Answers are grounded in this paper and cite their sources.
A new type of scintillation spectrometer is described. It is designed to suppress signals arising from incident photons that are not completely absorbed in the primary detecting NaI(Tl) crystal. When an undesired signal is detected by means of a cluster of NaI(Tl) crystals surrounding the primary crystal, it is removed through the discriminating action of an anticoincidence electronic gate. Under conditions of good discrimination, this type of arrangement offers the advantage of yielding the gamma-ray spectrum directly without any interpretation of results which might become laborious, particularly if the incident gamma radiation has a complex spectrum. Results are presented that illustrate the discrimination action of the device by means of photographic representations of the scintillation gamma-ray spectra of Cs137 and Na22.
No takes yet. Share an insight, caveat, or question.
Richard D. Albert (1953) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: