Certain lanthanide-based solid-state materials provide highly efficient luminescent detectors. This article presents a study of a new radiation detector that can be easily produced from high-purity precursors of calcium fluoride and ytterbium and thulium oxides. A first SEM/EDX analysis demonstrates the composition and distribution of the activators that were prepared in the samples. The thermoluminescence study of CaF 2 :Yb,Tm showed that peak-II at 200 °C (fading of 4%) is markedly reduced respect to CaF 2 :Tm at 220 °C (fading of 8%). Unlike the reference sample CaF-Tm (CaF 2 :Tm), the new CaF 2 :Yb,Tm material (sample CaF-Yb-Tm) showed a stable low temperature TL signal (165 °C) at peak-I, as well as a high temperature peak-III at 315 °C that is quite stable (fading of 12%). A thermal lag effect was observed around -20 °C for peaks II and III by reducing the heating rate from 5 °C/s to 2 °C/s, accompanied by a 35% (peak-I) and 17% (peak-II) increase in the integrated TL intensities. Dose variation from 0.17 to 3.5 Gy provided first-order kinetics whose parameters were better fitted to 8 peaks than to 7 peaks, reporting linearity for all the peaks (R 2 > 0.991) over the entire dose range studied. A relevant preliminary finding of this study is that the stability of this multi-peak TL material, CaF 2 :Yb,Tm, makes it more reliable than CaF 2 :Tm for radiation dosimetry applications.