Answers are grounded in this paper and cite their sources.
Elliptical crystal spectrometers equipped with time-gated microchannel plate (MCP) detectors provide time-, space-, and spectrally resolved data. A common problem is that the number of time resolution elements is limited by the number of MCP frames. The number of frames that fit on a given MCP is limited by the image size and the alignment tolerance. At the Z facility these problems have been addressed with twin-elliptical-crystal spectrometers. Using two crystals and detectors doubles the number of frames available. This enables measurements with ∼350ps time resolution while still recording data from an ∼4ns wide time window. Alternatively, the twin crystal design allows simultaneous measurements with different crystals to investigate different spectral regimes.
No takes yet. Share an insight, caveat, or question.
Lake et al. (2006) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: