Key result
CRT-D reprogramming fails to detect recurrent VF after MI-induced hyposensing, requiring surgical electrode replacement.
Why the study?
Does surgical replacement of the pacing/sensing electrode resolve CRT-D hyposensing caused by acute myocardial infarction when reprogramming fails?
Case Report (n=1)
Does surgical replacement of the pacing/sensing electrode resolve CRT-D hyposensing caused by acute myocardial infarction when reprogramming fails?
In cases of critically low spontaneous R-wave hyposensing due to acute myocardial infarction, reprogramming may be insufficient, necessitating surgical replacement of the pacing and sensing electrode.
Answers are grounded in this paper and cite their sources.
May signal acute ischemia via remote R-wave monitoring in CRT-D patients; leaves open whether parameter alerts improve outcomes.
The article describes a clinical case of a patient with a previously implanted cardiac resynchronization therapy with defibrillator function (CRT-D). For a long time, the patient was observed in the implanting clinic and the data was also transmitted via remote monitoring. The developing acute myocardial infarction in the right coronary artery caused a change in the parameter of the amplitude of the spontaneous R-wave in the patient. Subsequent ventricular fibrillation (VF) due to stent thrombosis was undetected by the defibrillator and resulted in no detection and cessation of VF, despite the implanted CRT-D. Changing the programming parameters did not affect the possibility of detecting a recurrent VF. The VF that occurred twice was stopped by the first discharge of the external device. In view of the remaining low hyposensing, a decision was made to reimplant the detecting part of the right ventricular electrode, after which defibrillation test proved to be effective for termination induced VF. We believe that at critically low parameters of spontaneous R-wave hyposensing, for the device to detect low-amplitude ventricular arrhythmias, programming the parameters does not solve the problem of hyposensing. Current myocardial infarction as the cause of hyposensing may require active surgical tactics replacement of the pacing and sensing electrode, in cases where the device connector allows this.
No takes yet. Share an insight, caveat, or question.
Postol et al. (2023) conducted a case report in Acute myocardial infarction with CRT-D and undetected ventricular fibrillation (n=1). Surgical replacement of the pacing and sensing electrode was evaluated on Detection and termination of ventricular fibrillation. In a patient with CRT-D and acute myocardial infarction causing R-wave hyposensing, device reprogramming failed to detect recurrent VF, necessitating surgical replacement of the sensing electrode.
Synapse has enriched 10 related papers. Consider them for comparative context: