Key result
4D Doppler shows good agreement with conventional ultrasound for assessing carotid stenosis via ICA PSV.
Why the study?
Does a 4-dimensional semiautomated Doppler system accurately assess carotid stenosis compared to conventional Doppler ultrasound in patients referred for clinical carotid ultrasound?
Observational (n=70)
Does a 4-dimensional semiautomated Doppler system accurately assess carotid stenosis compared to conventional Doppler ultrasound in patients referred for clinical carotid ultrasound?
Effect estimate: ICC 0.83
p-value: p=< .0001
A 4-dimensional semiautomated Doppler device provides comparable assessment of carotid stenosis to conventional Doppler ultrasound.
May support 4D Doppler as comparable for carotid stenosis grading; hypothesis-generating and should not yet change practice.
OBJECTIVE: The purpose of this study was to compare peak systolic velocities (PSVs) and the degree of stenosis obtained with a real-time 3-dimensional (ie, 4-dimensional) Doppler ultrasound scanner (Encore PV; VueSonix Sensors Inc, Wayne, PA) to conventional Doppler ultrasound imaging of the carotid arteries (common [CCA], internal [ICA], and external [ECA]). A secondary goal was to assess Encore volume flow measurements. METHODS: Seventy patients referred for clinical carotid ultrasound participated in this pilot study. Peak systolic velocities of the CCA, ECA, and ICA were obtained bilaterally. The degree of stenosis in the ICA was calculated based on the ICA PSV and ICA/CCA PSV ratio. The Encore detects all 3-dimensional blood flow velocity vectors within 10-s longitudinal volumes of the ICA, ECA, and CCA. On the Encore, a reader determined the centerline of the vessels. The PSV and volume flow were then automatically calculated. The flow measurement error was obtained by comparing the CCA flow to the ICA and ECA flow. Data were compared using linear regression, intraclass correlation coefficients (ICCs), and Bland-Altman analysis. RESULTS: Due to technical difficulties, only 59 patients (323 vessel segments) were available for analysis. There was good agreement between methods for assessing the degree of stenosis based on the ICA PSV (ICC = 0.83; P < .0001) and, to a lesser degree, on the ICA/CCA PSV ratio (ICC = 0.65; P < .0001). Peak systolic velocity measurements obtained with conventional ultrasound and the Encore correlated in all vessels (r >or= 0.32; P < .002), and Bland-Altman analysis showed reasonable variations. The Encore mean volume flow error +/- SD was -4.1% +/- 66.4% and was not biased (P = .57). CONCLUSIONS: A new semiautomated 4-dimensional Doppler device is comparable to conventional Doppler ultrasound for assessment of carotid stenosis.
No takes yet. Share an insight, caveat, or question.
Forsberg et al. (2008) conducted an observational in Carotid stenosis (n=70). 4-dimensional Doppler ultrasound scanner (Encore PV) vs. Conventional Doppler ultrasound imaging was evaluated on Degree of stenosis based on the internal carotid artery peak systolic velocity (ICC 0.83, p=< .0001). A 4-dimensional Doppler ultrasound scanner showed good agreement with conventional ultrasound for assessing carotid stenosis based on ICA peak systolic velocity (ICC = 0.83; P < .0001).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: