Answers are grounded in this paper and cite their sources.
Procedures are described for insulating metal scanning tunneling microscope (STM) tips with either glass or polymer coatings. In solutions containing 0.10 M of a reversible redox couple, Fe(CN)−3/−46, the faradaic limiting current to polymer coated tips was 200–500 pA and that for glass coated tips was <10 pA. For polymer insulated tips, steady-state currents of 10–100 pA were observed at tip-sample displacements less than 0.3 μm. The suppression of faradaic current achieved by these coating procedures enabled the collection of the first atomic resolution STM images of highly ordered pyrolytic graphite electrodes in contact with redox-active electrolytes. Preliminary data for the in situ electrochemical characterization of these tips are also discussed.
No takes yet. Share an insight, caveat, or question.
Heben et al. (1989) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: