Key result
Contactless millimeter-wave FMCW radar detects cardiac systole and diastole with close correlation to simultaneous ECG.
Why the study?
The study aimed to detect cardiac movement non-invasively using millimeter-wave FMCW radar and correlate it with electrocardiogram reference signals.
Can millimeter-wave FMCW radar accurately detect cardiac mechanical phases compared to standard ECG?
Observational (n=20)
No
Can millimeter-wave FMCW radar accurately detect cardiac mechanical phases compared to standard ECG?
Contactless millimeter-wave FMCW radar can accurately detect cardiac mechanical phases (systole and diastole) that correlate with electrical signals from an ECG.
Supports contactless cardiac monitoring feasibility; hypothesis-generating and requires clinical validation before practice adoption.
The work presented in this article aims to detect the cardiac movement of a person in a non-invasive way and to correlate it with a reference signal in the medical field: the electrocardiogram. To achieve this goal, a measurement campaign was carried out on twenty consenting individuals. On the one hand, the mechanical signal, the movement of the subject’s chest induced by the heartbeat, is recorded via a FMCW radar and on the other hand, the electrical signal of the heart is recorded via an ECG acquisition board. Signal processing functions and different filtering will allow the correlation of the radar and ECG signals. This study is conducted on apnea recordings in order to remove the impact of breathing on the movement of the chest. When the subject holds his or her breath, the two important phases of cardiac movement via radar capture can be detected: the systole and the diastole. The delay between the mechanical signal of the heart and the electrical signal of the heart, already explained by medicine, is well noted. The accuracy of motion detection provided by the radar allowed us to highlight the reproducibility of the chest movements detected during a capture. Their correlation with ECG data validates the proposed hypotheses.
No takes yet. Share an insight, caveat, or question.
Brulc et al. (2023) conducted an observational in Healthy individuals (n=20). Millimeter-wave FMCW radar vs. Electrocardiogram (ECG) was evaluated on Detection and reproducibility of cardiac movement (systole and diastole). Contactless millimeter-wave FMCW radar successfully detected chest movements corresponding to cardiac systole and diastole, correlating closely with simultaneous ECG recordings.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: