Answers are grounded in this paper and cite their sources.
Experimental study demonstrates accurate tumor depth estimation in tissue phantoms and porcine liver, highlighting its potential for robot-assisted surgery.
Robot-assisted minimally invasive surgery lacks haptic feedback for localizing subsurface tumors and estimating depth. We present a compact, high-density fiber Bragg grating (FBG) tactile sensor array for robotic palpation. A hexagonal close-packed topology increases spatial sampling within an 8.5-mm maximum lateral span while maintaining high sensitivity. Static and dynamic tests demonstrate linear force-wavelength responses across seven channels, a worst-case nominal interrogator-limited force resolution of 8.43 mN, and <1% full-scale dynamic error. We further introduce FBG-PatchFormer, which uses contrastive self-supervised pretraining on unlabeled palpation windows followed by supervised finetuning for depth classification. In phantoms with embedded inclusions, the model achieved 99.60% record-level accuracy for ten classes (blank and 2–10 mm) within the calibrated target-stiffness regime. Cross-condition tests showed transfer across inclusion shape and moderate surface curvature; performance under background-stiffness shift was partially recovered using three labeled records per class. In vivo porcine-liver experiments demonstrated force-map-based localization under physiological motion and fluid interference, supporting the feasibility of the sensing system for robotic palpation.
No takes yet. Share an insight, caveat, or question.
Dong et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: