Cadaver feasibility study demonstrates reliable detection of ventricular anatomy using 0.05 T MRI in dogs, suggesting utility for hydrocephalus screening.
Magnetic resonance imaging (MRI) systems continue to grow in number in veterinary and human medicine with higher field strength magnets increasingly available; however, the associated high costs have created healthcare inequities. This technical feasibility study aimed to evaluate a low-cost, portable, and permanent magnet-based 0.05 T low-field system (LF), originally designed for human neuroimaging, for assessing the canine brain. By using 23 canine cadavers, a scanning protocol was developed, consisting of transverse T1-weighted (T1w) and T2-weighted (T2w) sequences, and compared to a 1.5 T (HF) protocol for the visibility of 22 anatomical features. For the purpose of hydrocephalus evaluation, the LF system showed non-inferior detection of the lateral ventricles (LV) and mesencephalic aqueduct on T2w sequences with >90% conditional probability for detecting these features on the LF, given they were present on the HF. For T1w sequences, the LF system had inconclusive non-inferiority for detection of LV with 76.6% conditional probability. In summary, a canine brain scanning protocol was successfully developed for a 0.05 T MRI with preliminary evaluation suggestive that this device can aid with the detection of hydrocephalus. Future in vivo studies are required to evaluate the device's true clinical application.
No takes yet. Share an insight, caveat, or question.
Burgers et al. (2026) studied this question.