Key result
Escalating AED shocks match non-escalating shocks for ROSC and myocardial damage in a porcine model.
Why the study?
AEDs use various shock protocols with different characteristics when deployed in pediatric mode, prompting an assessment and comparison of their safety and efficacy.
Does an escalating pediatric defibrillation protocol compared to a non-escalating protocol affect myocardial damage or defibrillation efficacy in a porcine model?
RCT (n=22)
randomized
Does an escalating pediatric defibrillation protocol compared to a non-escalating protocol affect myocardial damage or defibrillation efficacy in a porcine model?
Typical variations in pediatric AED shock protocols (escalating vs non-escalating) do not cause significant myocardial damage and are equally effective in a porcine model.
Answers are grounded in this paper and cite their sources.
Reassures on protocol equivalence in porcine model; leaves optimal pediatric AED strategy unresolved pending clinical data.
AIM: Automated external defibrillators (AEDs) use various shock protocols with different characteristics when deployed in pediatric mode. The aim of this study is to assess and compare the safety and efficacy of different AED pediatric protocols using novel experimental approaches. METHODS: Two defibrillation protocols (A and B) were assessed across two studies: Protocol A: escalating (50-75-90 J) defibrillation waveform with higher voltage, shorter duration and equal phase durations. Protocol B; non-escalating (50-50-50 J) defibrillation waveform with lower voltage, longer duration and unequal phase durations.Experiment 1: Isolated shock damage was assessed following shocks to 12 anesthetized pigs. Animals were randomized into two groups, receiving three shocks from Protocol A (50-75-90 J) or B (50-50-50 J). Cardiac function, cardiac troponin I (cTnI), creatine phosphokinase (CPK) and histopathology were analyzed. Experiment 2: Defibrillation safety and efficacy were assessed through shock success, ROSC, ST-segment deviation and contractility following 16 randomized shocks from protocol A or B delivered to 10 anesthetized pigs in VF. RESULTS: Experiment 1: No clinically meaningful difference in cTnI, CPK, ST-segment deviation, ejection fraction or histopathological damage was observed following defibrillation with either protocol. No difference was observed between protocols at any timepoint. Experiment 2: all defibrillation types demonstrated shock success and ROSC ≥ 97.5%. Post-ROSC contractility was similar between protocols. CONCLUSIONS: There is no evidence that administration of clinically relevant shock sequences, without experimental confounders, result in significant myocardial damage in this model of pediatric resuscitation. Typical variations in AED pediatric mode settings do not affect defibrillation safety and efficacy.
No takes yet. Share an insight, caveat, or question.
McCartney et al. (2022) conducted an RCT in Ventricular fibrillation / Pediatric resuscitation model (n=22). Protocol A (escalating defibrillation waveform) vs. Protocol B (non-escalating 50-50-50 J waveform) was evaluated on Myocardial damage (cTnI, CPK, ST-segment deviation, ejection fraction, histopathology) and shock success/ROSC. In a porcine model, escalating and non-escalating AED shock protocols both achieved ≥97.5% shock success and ROSC without clinically meaningful differences in myocardial damage.