Key result
Commercial carotid shunts show ~2.8-fold variation in fluid delivery rates.
Why the study?
Do different commercially available carotid shunts have different flow rates when tested in a mock circuit?
Do different commercially available carotid shunts have different flow rates when tested in a mock circuit?
Commercially available carotid shunts exhibit significant hemodynamic differences in flow rates, challenging the assumption that all shunts provide equal cerebral protection during extracranial carotid surgery.
Answers are grounded in this paper and cite their sources.
Marked flow variation among carotid shunts warrants device-specific testing; leaves open effects on cerebral protection in carotid surgery.
A wide variety of carotid shunts are available for use in extracranial carotid surgery. Since it is commonly assumed that when properly positioned all shunts are equal in ability to protect the brain from cerebral ischemia, the choice of shunt is usually based on handling characteristics. However, after an intraoperative stroke occurred in a patient, we compared shunt flow rates using a simple and reproducible method of measurement. A mock circuit was created using a saline-filled fluid reservoir connected to the particular shunt being tested via 1/2-inch tubing. Hydrostatic pressure across the shunt was varied by changing the height of the reservoir, and the flow was collected over 30-second intervals. Multiple flow rate measurements were performed for each shunt with pressure gradients varying from 25 to 150 cm H2O. The data show significant hemodynamic differences among commercially available carotid shunts. A pressure gradient of 75 cm H2O produced a 2.8-fold variation in the amount of fluid delivered by various shunts. Minimal cerebral blood flow requirements and the possibility of underperfusion require that the surgeon consider such data in choosing an appropriate carotid shunt.
No takes yet. Share an insight, caveat, or question.
Grossi et al. (1993) studied extracranial carotid surgery. Commercially available carotid shunts vs. Various shunts was evaluated on Shunt flow rates (2.8-fold variation). Commercially available carotid shunts demonstrated significant hemodynamic differences, with a 2.8-fold variation in fluid delivery at a pressure gradient of 75 cm H2O.