Key result
Epicardial Coriolis sensor tracks LV dynamics paralleling LVdP/dtMAX, detecting a ~155% angular velocity increase with inotropes.
Why the study?
Can an epicardial gyroscopic sensor based on Coriolis force accurately measure cardiac apex rotation compared to LVdP/dtMAX in a sheep model?
Can an epicardial gyroscopic sensor based on Coriolis force accurately measure cardiac apex rotation compared to LVdP/dtMAX in a sheep model?
An implantable Coriolis force-based rotation sensor can effectively track LV torsional deformation in normal and failing hearts in a preclinical model.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for epicardial rotation sensors; clinical translation requires human validation before any practice change.
Cardiac apex rotation, quantified by sophisticated techniques (radiopaque markers and tagged magnetic resonance), has been shown to provide a sensitive index of left ventricle (LV) dynamics. The authors describe the first experimental assessment of cardiac apex rotation using a gyroscopic sensor based on Coriolis force, epicardially glued on the apex. Dynamics of apex rotation were evaluated in a sheep at baseline, after a positive inotropic drug infusion, and after impairment of cardiac function induced by coronary ligation. To evaluate the efficacy of the sensor to monitor cardiac function, results were compared to contractility variations expressed by the maximum value of the first derivative of LV pressure (LVdP/dtMAX). After inotropic drug infusion, a parallel increasing trend resulted for LVdP/dtMAX, for the maximum value of angular velocity measured by the sensor, and for apex rotation angle derived from velocity signal (+146%, +155%, and +11% from baseline, respectively), whereas a decreasing trend of all three parameters resulted after coronary ligation (-35%, -31%, and -65%). The twist pattern also was altered from baseline. These initial results suggest that the use of an implantable rotation sensor based on Coriolis force can be an efficient and effective tool to assess LV torsional deformation both in normal and failing hearts.
No takes yet. Share an insight, caveat, or question.
Marcelli et al. (2005) studied Cardiac function assessment (n=1). Epicardial Coriolis force sensor vs. Baseline and LVdP/dtMAX was evaluated on Maximum value of angular velocity and apex rotation angle. An epicardial Coriolis force sensor tracked LV dynamics, with maximum angular velocity increasing 155% after inotropic infusion and decreasing 31% after coronary ligation, paralleling LVdP/dtMAX.