Key result
Monopolar electrocautery shows 0 device failures at one month in patients with hypoglossal nerve stimulators.
Why the study?
Monopolar electrocautery is traditionally avoided with implanted hardware due to concerns of electromagnetic interference, but no study existed on its impact in patients with hypoglossal nerve stimulators.
Does the use of monopolar electrocautery cause device failure or malfunction in patients with implanted hypoglossal nerve stimulators?
Observational (n=85)
No
Does the use of monopolar electrocautery cause device failure or malfunction in patients with implanted hypoglossal nerve stimulators?
Monopolar electrocautery appears safe for use outside the immediate vicinity of implanted hypoglossal nerve stimulators without compromising device function.
Answers are grounded in this paper and cite their sources.
May support monopolar electrocautery use away from hypoglossal nerve stimulators; hypothesis-generating and requires prospective confirmation.
OBJECTIVES: Monopolar electrocautery (MPE) is traditionally avoided in patients with implanted hardware due to concerns about electromagnetic interference. While MPE use in cardiac, spinal cord, and cochlear implants has been documented, no study exists on its impact on hypoglossal nerve stimulators (HGNS). Here we evaluate the impact of MPE use in patients with HGNS implants. METHODS: This is a retrospective review of 85 consecutive HGNS cases at a tertiary academic center from October 2012 to October 2023. Post-HGNS implantation surgical procedures were categorized into surgeries with MPE (n = 38) and without MPE (n = 47). Data on demographics, polysomnography, and device outcomes were analyzed using descriptive statistics, Wilcoxon signed-rank tests, and chi-squared tests. RESULTS: Of the 85 patients, 38 (45%) underwent MPE-involved surgeries, with 7 (18%) head and neck, 20 (53%) torso and upper extremities, and 11 (29%) below the torso. No significant difference in outcomes was found between surgeries above or below the torso (p = 0.478). Up to 1 month post-MPE exposure, no device failure or malfunction occurred. Among 11 patients with sleep studies before and after MPE exposure, there was no significant difference in AHI (p = 1.00). CONCLUSION: This is the first study to assess the safety of MPE in patients with HGNS implants, suggesting that MPE use outside the immediate vicinity of HGNS implants is safe and does not compromise HGNS function.
No takes yet. Share an insight, caveat, or question.
Abdul‐Rahman et al. (2025) conducted an observational in Implanted hypoglossal nerve stimulators (n=85). Monopolar electrocautery (MPE) vs. Surgeries without MPE was evaluated on Device failure or malfunction. Monopolar electrocautery use in patients with hypoglossal nerve stimulators resulted in 0 device failures or malfunctions up to 1 month post-exposure.